Title Accuracy of SRS Dose Delivery Using the TomoTherapy Hi-Art System
Degree Master of Science (M.S.)
Department Physics & Astronomy
Advisory Committee
Advisor Name Title
Parker, Brent C. Committee Chair
Fields, Robert Committee Member
Gonzalez,Gabriela Committee Member
Hogstrom, Kenneth Committee Member
Matthews, Kenneth Committee Member
Price, Michael Committee Member
Keywords
  • radiosurgery
  • Tomotherapy
  • SRS
  • stereotactic
Date of Defense 2010-07-15
Availability unrestricted
Abstract
Purpose: To quantify the accuracy and precision of both target positioning and dose delivery for intracranial radiosurgery delivered with the TomoTherapy Hi-Art System using a non-invasive immobilization device.

Methods: Techniques developed by Vinci et al (2007) were refined for the measurement of dose distributions in each principal plane using a CIRS head phantom. Pieces of Gafchromic EBT2 film were cut and digitized using a template developed by Vinci et al (2007). A plan was created for a 2 cm diameter x 2 cm long cylindrical target in the TomoTherapy treatment planning system (TPS) version 3.2.1. Intentional misalignments of 5 mm in each of the principal directions were applied to the phantom prior to treatment delivery. The MVCT feature of the TomoTherapy Hi-Art system was used to correct for these misalignments, and then the treatment was delivered. Measured dose distributions (film) were registered to the calculated dose distributions (TPS planar dose) and compared.

Results: Alignment errors (displacement between the midpoints of the measured and calculated 70% dose points; mean ± standard deviation) were -0.15 ± 0.47 mm (range: -1.97 to 0.8 mm), -0.36 ± 0.56 mm (range: -1.25 to 0.63 mm), and -0.67 ± 0.93 mm (range: -3.04 to 0.90 mm) in the superior-inferior, anterior-posterior, and lateral directions, respectively. Positional errors of the 80% dose points in millimeters were 1.28 ± 0.91 (range: -0.09 to 3.62), -0.02 ± 0.96 (range: -2.24 to 1.72), -0.04 ± 0.62 (range: -1.24 to 1.25), 0.64 ± 0.52 (range: -0.35 to 1.55), 0.30 ± 0.52 (range: -1.57 to 1.28), and 0.60 ± 0.46 (range: -0.26 to 2.39) for the right, left, posterior, anterior, inferior, and superior directions, respectively.

Conclusions: Using a non-invasive immobilization device, 1.98 mm dose voxel size, and manual lateral couch positioning, the spatial accuracy of dose delivery with the TomoTherapy Hi•Art System was not within 1 mm as hypothesized.

Files
  Filename       Size       Approximate Download Time (Hours:Minutes:Seconds) 
 
 28.8 Modem   56K Modem   ISDN (64 Kb)   ISDN (128 Kb)   Higher-speed Access 
  BATTEthesis.pdf 1.16 Mb 00:05:22 00:02:45 00:02:25 00:01:12 00:00:06

Browse All Available ETDs by ( Author | Department )

If you have questions or technical problems, please Contact LSU-ETD Support.