Title Evaluation of a Desktop Computed Radiography System for IMRT Dosimetry
Degree Master of Science (M.S.)
Department Physics & Astronomy
Advisory Committee
Advisor Name Title
Oscar Hidalgo-Salvatierra Committee Chair
James M. Matthews Committee Co-Chair
Erno Sajo Committee Member
Kenneth Matthews II Committee Member
Sheldon A. Johnson Committee Member
Keywords
  • intensity profile
  • fluence map
  • dose comparison
Date of Defense 2004-08-23
Availability unrestricted
Abstract
Different techniques have been developed and used to evaluate dose distribution calculation accuracy and dose delivery reproducibility as a part of patient-specific IMRT QA – e.q. film dosimetry, ionization chambers, and diode arrays. To verify that the calculated dose distribution is delivered accurately during treatment, film dosimetry is usually used. The accuracy and reproducibility of film optical density as an indicator of dose is influenced by several variables, including the chemical processing and scanning conditions. This study investigates the possibility to use a desktop computed radiography (CR) system for patient-specific intensity modulated radiation therapy (IMRT) quality assurance (QA).

A study was done at Mary Bird Perkins Cancer Center, Baton Rouge, LA; where phantom IMRT plans are calculated using an ADAC Pinnacle3 treatment planning system. A Kodak ACR-2000i system is used for the study together with Kodak flexible phosphor screens (plates).

In this study, 778 CR plate exposures were done. Several tests were performed including evaluation of the CR plate response dependency when exposed to changes in either setup or scan conditions.

Calibration curves were generated for three different energies: 4 MV, 6 MV and 10 MV. Using these calibration curves, the CR plates’ response and behavior as an IMRT tool was analyzed using 10 different patients’ IMRT plans for each energy with approximately 7 fields per patient. Analysis of film was done with commercial IMRT analysis software. Analysis of CR plate data was done in IDL (Research Systems, Inc.), with programs written in house, and included several separate algorithms including automatic image registration. This algorithm uses the Fourier-Mellin transform for automatic image registration.

It was found that CR plates showed generally good agreement with the planned values with some significant over-response in the low dose regions, which can be reduced by filtration and improved calibration curves.

In view of the results presented, a CR system stands as a potentially fast and practical tool for IMRT patient-specific treatment QA.

Files
  Filename       Size       Approximate Download Time (Hours:Minutes:Seconds) 
 
 28.8 Modem   56K Modem   ISDN (64 Kb)   ISDN (128 Kb)   Higher-speed Access 
  Jurkovic_thesis.pdf 4.23 Mb 00:19:35 00:10:04 00:08:48 00:04:24 00:00:22

Browse All Available ETDs by ( Author | Department )

If you have questions or technical problems, please Contact LSU-ETD Support.